General Introduction

Yokogawa GD40 is the detector of GD40/GD402 gas density analyzer which applicable for intrinsically safe and explosion-proof zones, it is designed to be virtually maintenance free for all accepted applications which owns several outstanding features as : resistant to external vibrations; outstanding stability against sudden changes in gas temperature; multi-mode self-oscillation circuit minimizes drift caused by the sensor itself or by oil mist, dust, moisture, etc. sticking to the sensor; easy cleaning and regeneration of sensor which can be easily cleaned and returned to its original condition when the sensor be contaminated with dust and mist; only routine maintenance is required, for example, once per 3 months depending on application.


General Specification

Main Characters

Description

Material exposed to gas

SUS316 stainless steel, Acrylonitrile   Butadiene Rubber and Fluorine-contained Rubber (o-ring)

Ambient conditions

Temperature: -10 to 60°C (14 to 140°F), Humidity:   5 to 95%RH

Installation

Pipe-mounted or on panel

Ingress Protection

IP65

Construction

GD40G gas density analyzer: General   purpose detector for Non-Explosion-proof zones

GD40T gas density analyzer: FM   Explosion-proof and Intrinsically safe Approval

GD40V gas density analyzer: CSA Explosion-proof   and Intrinsically safe Approval

GD40R gas density analyzer: TIIS   Explosion-proof and Intrinsically safe Approval

3.Model Selection of GD40G GD40R

Model

Suffix Codes

Description

GD40G



General purpose detector. 1/4 NPT gas threads and 1/2   NPT gland threads. No cable gland included. Mounting hardware included

GD40R



TIIS certified explosion proof detector. Rc1/4 gas   threads and G1/2 gland threads. Cable gland included. Mounting hardware   included.

Label approval

-E


English   label, no approval (only GD40G)

-J


TIIS   approval, English label (only GD40R)

Options


/EJAJ1

TIIS certified EJX mounted with detector on mounting   plate. Rc1/4 gas threads and G1/2 gland thread. Cable gland included.(only   GD40R)


/EJAJ1T

TIIS certified EJX mounted with detector on mounting   plate. Rc1/4 gas threads and G1/2 gland thread. Cable gland included. EJX   with TAG (only GD40R)


/EJAF2

EJX mounted with detector on mounting plate.1/4 NPT gas   threads and 1/2 NPT gland threads. No cable gland included.(only GD40G)


/EJAF2T

EJX mounted with detector on mounting plate.1/4 NPT gas   threads and 1/2 NPT gland threads. No cable gland included. EJX with TAG   (only GD40G)


4.Model Selection of GD40T GD40V

Model

Suffix Codes

Description

GD40T


FM certified explosion proof detector. 1/4 NPT gas   threads and 1/2 NPT gland thread. No cable gland included. Mounting hardware   included

GD40V


CSA certified explosion proof detector. 1/4 NPT gas   threads and 1/2 NPT gland thread. No cable gland included. Mounting hardware   included.

Options

/EJAF3

FM certified EJX mounted with detector on mounting   plate.1/4 NPT gas threads and 1/2 NPT gland thread. No cable gland included.   (only GD40T)

/EJAF3T

FM certified EJX mounted with detector on mounting plate.1/4NPT   gas threads and 1/2 NPT gland thread. No cable gland included. EJX with TAG   (only GD40T)

/EJAF4

CSA certified EJX mounted with detector on mounting   plate. 1/4 NPT gas threads and 1/2 NPT gland thread. No cable gland included.   (only GD40V)

/EJAF4T

CSA certified EJX mounted with detector on mounting   plate. 1/4 NPT gas threads and 1/2 NPT gland thread. No cable gland included.   EJX with TAG (only GD40V)